CCMI Leica SP2 AOBS DIC Quick Ref
- Establish Kohler illumination:
- Set scope to USER (white light).
- Ensure Vis/Scan mirror switch on condenser turret is set to Vis.
- Adjust white light intensity to preference.
- Focus on specimen (you may need to switch back to UV to do this).
- Close condenser field iris (move upper lever to left).
- Open condenser aperture iris (move lower lever to right).
- Focus image of field iris with condenser focus control.
- Open field iris until image is almost the same size as your field of view.
- Center image of field iris with condenser centering controls.
- Open field iris until image moves out of field of view.
- Close aperture iris about half way.
- Ensure upper and lower polarizing filters are inserted.
- Make fine adjustments to the upper polarizer to get darkest image.
- Set prism on condenser turret (upper) to match the objective in use.
- NB: 20x/0.7 IMM uses the 40x prism.
- Set prism on objective turret (lower) to C, D or E depending on objective in use.
- 20x/0.7 IMM and 40x/0.85 use C.
- 63x/1.2 Water uses D.
- 63x/1.4 Oil uses E.
- Make fine adjustments to DIC prism position to optimize contrast.
- Set mirror above condenser to SCAN (care - turn SLOWLY!).
- Set lower polarizer to the OUT position
- Activate transmitted light detector in software.
- Set scan control to CONTINUOUS.
- Make fine adjustments to DIC prism position, upper polarizer and detector controls for best image.
Note: For best fluorescence resolution it is best to proceed as follows:
- Acquire fluorescent image with DIC prism set to BF.
- Turn off fluorescence detectors and turn on transmitted detector.
- Acquire DIC image with DIC prism set to C, D or E.
- Overlay images in software (<Process>->Editing->Merging - select Ch, check Append box.)
jdc 0409; Last modified 041011